User login

Maintenance for Varian979

Date: 
Tuesday, January 23, 2007
Synopsis: 
maintaining spectrometer tube assembly
Equipment/Beam Line: 
Description: 

Hour meter=913 hrs Spectrometer tube assembly including Ion source, button TC assembly, and per amplifier were taken apart and serviced. Few o-rings were replaced with buna. Ground slit plate was a bit discolored. It was polished by Scotch-Brite pad carefully. After assembly system tested and was functioning normally. See the picture.

Author: 
Rahim Abasalti