This pump and DRP-15 were sent for STANDARD OVERHAUL. Please see attachment for repair quotation.
Hour meter: 6975 hrs
Requisition number for repair: 1022841
Disassembly Observations:
Pump does not start. > > Frequency converter is in good condition. > > Pump doesn't spin by hand. > > Oil is a little dark and may have some moisture. > > Oil found in all stages of the pump. > > Lip seals are in bad condition. > > Gears are in good condition. > > Bearings will spin by hand. > > Driven shaft sized on the pass through hole on the LP1 and exhaust > stator. > > > > > > This pump also seems to sit for long periods of time due to the age of > the pump and the number of hours. We found oil in all stages of the > pump which is a indication that the seals failed. The pump should also > run for half an hour every couple of months when it is not used to > help keep the seals and bearings lubricated.
Attachment | Size |
---|---|
Lesker_ACP40_maintenance_SQ017129-1.pdf | 45.56 KB |